Troubleshooting

Frequency Response Plot

If a peak in the frequency response plot does not appear:

  1. Increase Sweep Width to the maximum value.
  2. Increase the Drive Amplitude in intervals of 200–300 mV until you have reached 2–3 V.
  3. If the peak still has not appeared, increase the Sweep Width by first increasing the Center Frequency, then maximizing the Sweep Width. If there is still no peak on the response plot, check the laser alignment.

Engaging the Sample

If the engage aborts because the tip is still too far away from the surface, return to Stage > Focus Surface and use the trackball to relocate the surface. After successful re-engaging, a well-formed image appears.

Cantilever Will Not Tune

If the Cantilever Tune fails, check the following:

 